The results of this in vitro/in silico study suggest that: i) Newer generations of THVs have ∼30% lower leaflet mechanical stress than the early generations; ii) For a given generation, SE THVs have lower leaflet mechanical stress than BE THVs. Further studies are needed to determine if these differences between new vs early THV generations and between SE vs BE THVs will translate into significant differences in long-term valve durability in vivo.