Physics and Imaging in Radiation Oncology · 2023 · 13 citations · 24 references
MedicineRadiation OncologyRadiologyBiomedical ImagingDiagnostic ImagingTreatment PlanningComputed TomographyComputed Tomography Dose ManagementExternal Beam RadiotherapyMedical ImagingRadiation TherapyRadiographic ImagingAdaptive Radiotherapy PlanningCt Image ReconstructionDose Distribution CalculationsImage-guided Dose AdaptationNeck Adaptive RadiotherapyPatient Radiation Dose OptimizationRadiation Dose OptimizationRadiographic Image Quality
Background and purpose: Accurate CT numbers in Cone Beam CT (CBCT) are crucial for precise dose calculations in adaptive radiotherapy (ART). This study aimed to generate synthetic CT (sCT) from CBCT using deep learning (DL) models in head and neck (HN) radiotherapy. Materials and methods: A novel DL model, the 'self-attention-residual-UNet' (ResUNet), was developed for accurate sCT generation. ResUNet incorporates a self-attention mechanism in its long skip connections to enhance information transfer between the encoder and decoder. Data from 93 HN patients, each with planning CT (pCT) and first-day CBCT images were used. Model performance was evaluated using two DL approaches (non-adversarial and adversarial training) and two model types (2D axial only vs. 2.5D axial, sagittal, and coronal). ResUNet was compared with the traditional UNet through image quality assessment (Mean Absolute Error (MAE), Peak-Signal-to-Noise Ratio (PSNR), Structural Similarity Index (SSIM)) and dose calculation accuracy evaluation (DVH deviation and gamma evaluation (1 %/1mm)). Results: Image similarity evaluation results for the 2.5D-ResUNet and 2.5D-UNet models were: MAE: 46±7 HU vs. 51±9 HU, PSNR: 66.6±2.0 dB vs. 65.8±1.8 dB, and SSIM: 0.81±0.04 vs. 0.79±0.05. There were no significant differences in dose calculation accuracy between DL models. Both models demonstrated DVH deviation below 0.5 % and a gamma-pass-rate (1 %/1mm) exceeding 97 %. Conclusions: ResUNet enhanced CT number accuracy and image quality of sCT and outperformed UNet in sCT generation from CBCT. This method holds promise for generating precise sCT for HN ART.
24
Deep Residual Learning for Image Recognition
Kaiming He, Xiangyu Zhang, Shaoqing Ren et al. · 2016 · 222.1K citations · Full text
Evaluation of the gamma dose distribution comparison method
Daniel A. Low, James F. Dempsey · Medical Physics · 2003 · 752 citations
Paired cycle‐GAN‐based image correction for quantitative cone‐beam computed tomography
Joseph Harms, Yang Lei, Tonghe Wang et al. · Medical Physics · 2019 · 255 citations · Full text
Synthetic CT generation from CBCT images via deep learning
Liyuan Chen, Xiao Liang, Chenyang Shen et al. · Medical Physics · 2019 · 198 citations · Full text