ACS Sensors · 2023 · 32 citations · 49 references
EngineeringBioengineeringChemistryBiosensorsElectrochemistrySensorsAptamer-based BiosensorsChemical SensorsElectrochemical BiosensorsElectrochemical Impedance SpectroscopyElectrochemical SensorsImpedance SpectroscopyAptamer BiosensorsAptamer-based SensorsConformational Change Upon BindingElectrochemical AptasensorsElectrochemical Molecular SensorsElectrochemical SensingElectron TransferTime Resolution
Electrochemical aptamer-based (EAB) sensors are capable of measuring the concentrations of specific molecules in vivo, in real time, and with a few-second time resolution. For their signal transduction mechanism, these sensors utilize a binding-induced conformational change in their target-recognizing, redox-reporter-modified aptamer to alter the rate of electron transfer between the reporter and the supporting electrode. While a variety of voltammetric techniques have been used to monitor this change in kinetics, they suffer from various drawbacks, including time resolution limited to several seconds and sensor-to-sensor variation that requires calibration to remove. Here, however, we show that the use of fast Fourier transform electrochemical impedance spectroscopy (FFT-EIS) to interrogate EAB sensors leads to improved (here better than 2 s) time resolution and calibration-free operation, even when such sensors are deployed in vivo. To showcase these benefits, we demonstrate the approach's ability to perform real-time molecular measurements in the veins of living rats.
49
Real surface area measurements in electrochemistry
S. Trasatti · Pure and Applied Chemistry · 1991 · 1.8K citations · Full text