Electrochemical-DNA Biosensor Development Based on a Modified Carbon Electrode with Gold Nanoparticles for Influenza A (H1N1) Detection: Effect of Spacer

Hyoeun Lee, Yun Ok Kang, Seong-Ho Choi

International Journal of Electrochemical Science · 2014 · 36 citations · 24 references

DOIFull text

Open access

Concepts

Abstract

An Electrochemical-DNA biosensor (E-DNA biosensor) was prepared successfully for electrochemical detection of a target DNA sequence, and the effective surface of the electrode was designed in order to achieve high sensitivity. First, electrode surfaces were modified chemically for immobilization of biomolecules by electrochemical reduction of 4-thiophenyl diazonium salt without any reducing agents. The thiol-modified electrode (TP-GCE) may be used to immobilize functionalized gold nanoparticles (Au NPs) and biomolecules. The characteristics of the modified electrode were obtained using cyclic voltammetry, contact angle, AFM, and XPS analysis. For confirmation of the effective influence of self-assembled layers (SAMs) with sensitivity of E-DNA biosensor, the spacer effect was determined by forming SAMs layer with 3-mercaptopropionic acid (MPA) and 11- mercaptoundecanoic acid (MUA) on the electrode. The E-DNA biosensor based on the MUA-modified electrode performed with higher sensitivity in hybridization between the probe DNA on the electrode and the target DNA than that based on the MPA-modified electrode because of the spacer effect that occurred owing to the long chemical chain of the MPA molecules. The new E-DNA biosensor is considered as the one of the most promising materials for highly sensitive clinical and other biotechnology applications.

References