Load fatigue performance of conical implant-abutment connections.
57
Citations
0
References
2011
Year
Ti conical abutments appear to have poorer load fatigue performance compared with earlier studies of external-hexagon connections. The load fatigue performance of Zr conical abutments varied and seemed to be highly system dependent. Many of the fractures in both the Ti and Zr abutment groups occurred within the implant, and retrieval would pose a significant clinical challenge. The clinician should weigh the mechanical, biologic, and esthetic considerations before selection of any implant system, connection type, or abutment material.