Biomedical Optics Express · 2023 · 18 citations · 42 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsBiophotonicsLabel-free ImagingMedical ImagingOptical CoherenceOptical Coherence TomographyOptical ImagingOptical TomographyBiological StructuresBiomedical MicroscopyLabel-free Cellular ImagingMetabolic Activity DetectionNon-invasive Imaging Assessment
Dynamic full-field optical coherence tomography (D-FFOCT) has recently emerged as an invaluable live label-free and non-invasive imaging modality able to image subcellular biological structures and their metabolic activity within complex 3D samples. However, D-FFOCT suffers from fringe artefacts when imaging near reflective surfaces and is highly sensitive to vibrations. Here, we present interface Self-Referenced (iSR) D-FFOCT, an alternative configuration to D-FFOCT that takes advantage of the presence of the sample coverslip in between the sample and the objective by using it as a defocused reference arm, thus avoiding the aforementioned artefacts. We demonstrate the ability of iSR D-FFOCT to image 2D fibroblast cell cultures, which are among the flattest mammalian cells.
42
David Huang, Eric A. Swanson, Charles P. Lin et al. · Science · 1991 · 13.7K citations
Quantitative phase imaging in biomedicine
YongKeun Park, Christian Depeursinge, Gabriel Popescu · Nature Photonics · 2018 · 1.8K citations · Full text