Effect of material type, torque value, and sterilization on linear displacements of a scan body: An in vitro study
27
Citations
18
References
2023
Year
Axial displacement of SBs was affected by material type, torque value, and sterilization. Ti SBs mostly had lower displacements than PEEK SBs. Application of 15 Ncm torque to tested PEEK SBs should be refrained from and a calibrated tightening tool may enable the application of 10 Ncm or lower torque values for lower displacements. Sterilization generally increased PEEK SB displacements.