Physics and Imaging in Radiation Oncology · 2022 · 26 citations · 12 references
MedicineCardiologyRadiation OncologyBiomedical ImagingRadiation TransportTreatment PlanningCardiac ImagingCardiothoracic ImagingImage-guided RadiotherapyRadiation TherapyStereotactic RadiosurgeryTarget Volume DelineationImage Registration Quality AssuranceRadiotherapy Dose PlanningRadiotherapy Target DelineationRegistrationVentricular Tachycardia AblationVentricular Tachycardia Mapping
A novel quality assurance process for electroanatomical mapping (EAM)-to-radiotherapy planning imaging (RTPI) target transport was assessed within the multi-center multi-platform framework of the RAdiosurgery for VENtricular TAchycardia (RAVENTA) trial. A stand-alone software (CARDIO-RT) was developed to enable platform independent registration of EAM and RTPI of the left ventricle (LV), based on pre-generated radiotherapy contours (RTC). LV-RTC were automatically segmented into the American-Heart-Association 17-segment-model and a manual 3D-3D method based on EAM 3D-geometry data and a semi-automated 2D-3D method based on EAM screenshot projections were developed. The quality of substrate transfer was evaluated in five clinical cases and the structural analyses showed substantial differences between manual target transfer and target transport using CARDIO-RT.
12
Oliver Blanck, Daniel Buergy, Maren Vens et al. · Clinical Research in Cardiology · 2020 · 71 citations · Full text
Judit Boda‐Heggemann, Oliver Blanck, Felix Mehrhof et al. · International Journal of Radiation Oncology*Biology*Physics · 2021 · 48 citations
Medicine, Cardiology, Surgery +13