The Effect of Paper on the Detection Limit of Paper-Based Potentiometric Chloride Sensors

Eliza J. Herrero, Blair K. Troudt, Philippe Bühlmann

Analytical Chemistry · 2022 · 10 citations · 40 references

Abstract

While paper is an excellent material for use in many other portable sensors, potentiometric paper-based sensors have been reported to perform worse than conventional rod-shaped electrodes, in particular in view of limits of detection (LODs). Reported here is an in-depth study of the lower LOD for Cl<sup>-</sup> measurements with paper-based devices comprising AgCl/Ag transducers. Contamination by Cl<sup>-</sup> from two commonly used device materials─a AgCl/Ag ink and so-called ashless filter paper─was found to increase the concentration of Cl<sup>-</sup> in paper-contained samples far above what is expected for the spontaneous dissolution of the transducer's AgCl, thereby worsening lower LODs. In addition, for the case of Ag<sup>+</sup>, the commonly hypothesized adsorption of metal cations onto filter paper was found not to significantly affect the performance of AgCl/Ag transducers. We note that in the context of chemical analysis, metal impurities of paper are often mentioned in the literature, but Cl<sup>-</sup> contamination of paper has been overlooked.

References

40