The Clinical Relevance of Instrumented Testing for ACL Insufficiency

Keith L. Markolf, Harlan C. Amstutz

Clinical Orthopaedics and Related Research · 1987 · 13 citations · 0 references

Concepts

Abstract

An instrumented clinical testing device developed at UCLA records a continuous anteroposterior force versus displacement curve of the tibia with respect to the femur at 20° of flexion. Laxity and stiffness are calculated from the response curve. With this device, 95% of normal knees have an anterior laxity less than 7.5 rnm and a side-to-side difference less than 2 mm. In contrast, an anterior cruciate ligament (ACL) absent knee has a mean anterior laxity of 10 mm and a mean side-to-side difference of 5 mm. In a small group of patients with an intraarticular ACL substitution using the medial or lateral one-third of the patellar tendon, laxity and stiffness of the injured knees were returned to within the normal range and remained constant three years after surgery. In a group of 76 patients treated with ACL substitution using the torn meniscus, 51% of the patients still had an anterior laxity outside the normal range 3.5 years after surgery. In a preliminary study of 19 patients receiving a Gore-Tex synthetic ACL substitution, 55% of the patients still had a side-to-side difference greater than 2 mm two years after the procedure. These studies illustrate the advantages of impartial, objective measurements of knee stability. Laxity and stiffness values can supplement, but never replace, a thorough patient examination and patient history. As sports medicine matures as a scientific discipline, improved instrumented test devices may ultimately provide a standardized means for reporting knee stability parameters.