Radiology · 1930 · 12 citations · 0 references
EngineeringElectrical EngineeringInstrumentationRadiation MeasurementRadiation ShieldingDetector SystemsIonization ChambersIonization TechniquesX-ray ShieldingAccuracyCurrent CompensationIonization Chamber ConstructionIonization Chamber DesignIonization Chamber PhysicsParallel-plate Ionization ChambersPrimary Standard Ionization ChambersSecondary Standard Ionization Chambers
A modification of the open-air parallel plate ionization chamber, readily portable and adapted to field use, is described. In this chamber, narrow guard plates are supplemented by a system of ten guard wires across the ends of the chamber. The wires are so spaced that X-ray scattering from them is negligible. By this means a uniform electrostatic field between electrodes is obtained without the use of wide guard plates and the large plate-to-shield spacing necessary in the primary standard. This design has permitted a decrease in the volume and weight of the secondary standard by a factor of 15. The results of direct comparisons between this chamber and the primary standard are given and indicate that the roentgen can be reproduced by the former with the same accuracy as the primary standard. The problem of diaphragming is identical for the two chambers, and for a given tube the same inverse power law holds for both. Like the standard chamber, the new chamber does not require that corrections be made for differences in X-ray quality. Any of the usual electrostatic measuring systems may be used with this chamber, although for precise work a null or current compensation system is desirable. It is expected that this type of chamber will be found useful in the international comparisons of the roentgen now pending, for which purpose the “fingerhutkammer” is generally inadequate.