Journal of Clinical Anesthesia · 2022 · 18 citations · 20 references
MedicinePerioperative MedicineSurgeryClinical Decision SupportPatient SafetyVascular SurgeryBlood Pressure ControlBlood Pressure ManagementCardiovascular AssessmentCardiovascular MonitoringHemodynamicsHypertension ManagementBlood Pressure EstimationBlood Pressure InterpretationDiastolic Blood Pressure EstimationHemodynamic Monitoring AlgorithmsMean Arterial Pressure EstimationNoninvasive Blood Pressure MeasurementObservational Study RegistrationSystolic Blood Pressure Estimation
STUDY OBJECTIVE: ). In this study we determined whether this new algorithm shows better agreement with invasive radial BP than the original one and whether in the operating room this algorithm can be used safely. DESIGN, SETTING AND PATIENTS: This observational study was conducted on thirty-three non-cardiac surgery patients. INTERVENTION AND MEASUREMENTS: were transformed. Agreement of systolic, mean, and diastolic arterial BP (SAP, MAP, and DAP, respectively) was assessed traditionally with Bland-Altman and trend analysis and clinically safety was quantified with error grid analyses. A bias (precision) of 5 (8) mmHg or less was considered adequate. MAIN RESULTS: (96 vs 77%, respectively). CONCLUSIONS: shows better agreement although clinical implications are small. This trial was registered with ClinicalTrials.gov (https://clinicaltrials.gov/ct2/show/NCT03795831).
20
Noninvasive Continuous Arterial Blood Pressure Monitoring with Nexfin®
Jerson R. Martina, Berend E. Westerhof, Jeroen van Goudoever et al. · Anesthesiology · 2012 · 291 citations
Medicine, Cardiology, Surgery +16