Midterm outcomes of AFX2 endografts used in combination with aortic cuffs

13

Citations

20

References

2022

Year

Abstract

Midterm outcomes of the AFX2 used in combination with an aortic cuff were acceptable, considering the rates of types I and III endoleaks. However, there were cases of sideways displacement that could cause future type IIIa endoleaks. When the AFX2 is used in combination with an aortic cuff, close surveillance for endograft deformations and subsequent adverse events, including type III endoleaks, is needed.