Heart Rhythm · 2022 · 48 citations · 16 references
MedicineCardiologyClinical CardiologyElectrophysiologyCardiac ArrhythmiasCardiac Conduction SystemCardiac PacingCardiovascular MonitoringElectrocardiographyVentricular ElectrophysiologyIntracardiac PacingLeadless PacemakersPacemaker ElectrocardiographyPhysiological Pacing
BACKGROUND: Previous studies demonstrated that accelerometer-based, mechanically timed atrioventricular synchrony (AVS) is feasible using a leadless ventricular pacemaker. OBJECTIVE: The purpose of this study was to determine the performance of a leadless ventricular pacemaker with accelerometer-based algorithms that provide AVS pacing. METHODS: AccelAV was a prospective, single-arm study to characterize AVS in patients implanted with a Micra AV, which uses the device accelerometer to mechanically detect atrial contractions and promote VDD pacing. The primary objective was to characterize resting AVS at 1 month in patients with complete atrioventricular block (AVB) and normal sinus function. RESULTS: A total of 152 patients (age 77 ± 11 years; 48% female) from 20 centers were enrolled and implanted with a leadless pacemaker. Among patients with normal sinus function and complete AVB (n = 54), mean resting AVS was 85.4% at 1 month, and ambulatory AVS was 74.8%. In the subset of patients (n = 20) with programming optimization, mean ambulatory AVS was 82.6%, representing a 10.5% improvement (P <.001). Quality of life as measured by the EQ-5D-3L (EuroQol Five-Dimensions Three-Level questionnaire) improved significantly from preimplant to 3 months (P = .031). In 37 patients with AVB at both 1 and 3 months, mean AVS during rest did not differ (86.1% vs 84.1%; P = .43). There were no upgrades to dual-chamber devices or cardiac resynchronization therapy through 3 months. CONCLUSION: Accelerometer-based mechanical atrial sensing provided by a leadless pacemaker implanted in the right ventricle significantly improves quality of life in a select cohort of patients with AV block and normal sinus function. AVS remained stable through 3 months, and there were no system upgrades to dual-chamber pacemakers.
16
US Valuation of the EQ-5D Health States
James W. Shaw, Jeffrey Johnson, Stephen Joel Coons · Medical Care · 2005 · 1.3K citations
Mikhael F. El‐Chami, Faisal Al-Samadi, Nicolas Clementy et al. · Heart Rhythm · 2018 · 349 citations · Full text
Medicine, Cardiology, Surgery +16