Journal of Hand Surgery (European Volume) · 2022 · 14 citations · 18 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryImplant SurgeryOrthopedicsProstheticsComplications Of Implant ProsthesesMetal ToxicityPeriprosthetic OsteolysisImplant Failure ManagementLong-term Implant Follow-upMetal Alloy ProsthesesMetallic ImplantsPeriprosthetic Bone ResorptionRadiographic Detection Of OsteolysisTitanium Materials
We analysed the relationship between serum metal ions, radiological periprosthetic osteolysis and the clinical features in a series of 25 patients treated with fourth-generation metal-on-polyethylene total wrist prostheses. The mean implant follow-up was 7 years. Our results show that titanium was the main elevated serum metal ion in patients with the prostheses that we used; elevated serum cobalt or chromium values were infrequent. The risk of loosening was higher in an implant older than 6 years, with more than five periprosthetic osteolysis points according to our radiograph zone system, and serum titanium values between 26 to 31 µg/L. The presence of metallosis pseudotumours does not guarantee but increases the risk of implant failure, which may be asymptomatic or associated with little pain. Level of evidence: IV
18