Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields

19

Citations

14

References

2022

Year

Abstract

The W2 demonstrated dosimetric characteristics that are suitable for radiosurgery field measurements. The detector agreed well with the RAZOR diode for output factors and scanned profiles and showed good agreement with the treatment planning system in measurements of clinical treatment plans.