Optics Express · 2022 · 47 citations · 52 references
Natural SciencesBiomedical EngineeringBiophysicsAcousticsBiophotonicsSensorsAbsorption SpectroscopyGas SensorsSpectral SensitivityFinite Element Acoustic SimulationInstrument Detection LimitMethod Detection LimitMultiple Reflection SuppressionPhotoacoustic Excitation SourcesPhotoacoustic Transducer DesignSignal-to-noise RatioSignal-to-noise Ratio Analysis
A high-sensitivity differential Helmholtz photoacoustic cell based on multiple reflection was reported, and its performance parameters and gas replacement time were optimized by finite element simulation. To realize the long absorption path of the measured gas, the collimated excitation light was reflected multiple times on the gold-plated wall of the absorption cavity, and the wavelength modulation technology was used to reduce the multiple reflection noise. Additionally, the differential could suppress external co-phase noise and double the photoacoustic signal. When a laser with a central wavelength of 1653 nm was employed as the excitation light source, the minimum detection limit of 177 ppb (signal-to-noise ratio, SNR = 1) for methane was achieved within a detection time of 1 s, and the corresponding normalized noise equivalent absorption coefficient was 4.1×10 –10 cm –1 WHZ –1/2 .
52
Quartz-tuning-fork enhanced photothermal spectroscopy for ultra-high sensitive trace gas detection
Yufei Ma, Ying He, Yao Tong et al. · Optics Express · 2018 · 345 citations · Full text