Frontiers in Physics · 2022 · 14 citations · 43 references
EngineeringMechanical EngineeringRadiologyAcousticsDiagnostic ImagingHeat TransferAcoustic ModelingMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsNoiseNoise PredictionNoise ReductionAcoustic NoiseAcoustic Noise ModelingAcoustic Noise ReductionCoil DesignCoolingMagnetic Field Gradient SystemsSound Pressure LevelTemperature Rise Analysis
Magnetic resonance imaging (MRI) gradient coils produce acoustic noise due to coil conductor vibrations caused by large Lorentz forces. Accurate sound pressure levels and modeling of heating are essential for the assessment of gradient coil safety. This work reviews the state-of-the-art numerical methods used in accurate gradient coil modeling and prediction of sound pressure levels (SPLs) and temperature rise. We review several approaches proposed for noise level reduction of high-performance gradient coils, with a maximum noise reduction of 20 decibels (dB) demonstrated. An efficient gradient cooling technique is also presented.
43
Markus Weiger, David O. Brunner, Benjamin E. Dietrich et al. · Magnetic Resonance in Medicine · 2013 · 170 citations · Full text
William A. Edelstein, Robert A. Hedeen, R. Mallozzi et al. · Magnetic Resonance Imaging · 2002 · 121 citations