Natural SciencesChemistryAnalytical ChemistryElectrochemistrySensorsBiological SensorsChemical SensorsElectrochemical BiosensorsElectrochemical DetectionElectrochemical SensorsIon-selective ElectrodesBovine Serum AlbuminElectrochemical SensingIon Selective SensorsIon-selective Electrode SensorsPolyanilinePotentiometric BiosensorsPotentiometric DetectionPotentiometric Sensors
We constructed a sensor for the determination of Fe2+ and/or Fe3+ ions that consists of a polyaniline layer as an ion-to-electron transducer; on top of it, chelating molecules are deposited (which can selectively chelate specific ions) and protected with a non-biofouling poly(2-methyl-2-oxazoline)s layer. We have shown that our potentiometric sensing layers show a rapid response to the presence of Fe2+ or Fe3+ ions, do not experience interference with other ions (such as Cu2+), and work in a biological environment in the presence of bovine serum albumin (as a model serum protein). The sensing layers detect iron ions in the concentration range from 5 nM to 50 µM.
35
Hassan Karimi‐Maleh, Samira Malekmohammadi, Nilofar Zakariae et al. · Journal of Molecular Liquids · 2020 · 283 citations