Publication | Open Access
Detection of rotational errors in single‐isocenter multiple‐target radiosurgery: Is a routine off‐axis Winston–Lutz test necessary?
23
Citations
29
References
2022
Year
It is recommended that off-axis WL testing be conducted regularly to supplement routine monthly machine QA, as it is sensitive to errors that patient-specific QA may not detect. This frequency should be determined by individual departments, with consideration of GTV-PTV margins used, limitations on target off-axis distances, and routine mechanical QA results for particular linacs.