EngineeringBiotechnologyMolecular BiologyAnalytical ChemistryBiosensorsElectrochemistryDna BiosensorsElectrochemical BiosensorsElectrochemical DetectionElectrochemical Impedance SpectroscopyNucleic Acid BiosensorsPolymerase Chain ReactionElectrochemical Impedance MonitoringImpedance MeasurementOptical DetectionPolymerase Chain Reaction (Pcr) AmplificationPolymerase Chain Reaction (Pcr) AssaysPolymerase Chain Reaction BiosensorsReal-time Polymerase Chain Reaction (Pcr)Real-time Polymerase Chain Reaction Assays
ions, which could be detected as a steep Rct, increased at specific heat cycles depending on the amount of template DNA. The Rct increase per heat cycle, ΔRct, showed a peak at the same heat cycle as optical detection, proving that PCR can be accurately monitored in real time by impedance measurement. This simple method will enable a cost-effective and portable PCR device.