Journal of Oral Science · 2022 · 14 citations · 23 references
DentistryImplant DentistryOrthodonticsImplant ImagingImplant SurgerySurgical OrthodonticsBiomechanics Of AnchorageRoot Proximity AssessmentContact Length MeasurementImplant Failure MechanismsLogistic Regression AnalysisMini-implant Failure RiskMini-implant Root ContactOrthodontic Mini-implantsRoot Proximity AvoidanceRoot-implant Proximity
PURPOSE: This study aimed to evaluate the causal relationship between root contact and root contact length of orthodontic miniscrew implants (miniscrews), and assessed factors for implant failure through cone-beam computed tomography (CBCT) images. METHODS: First, a logistic regression analysis was performed to investigate the most relevant factors for implant failure based on sex, age, vertical skeletal pattern, horizontal skeletal pattern, and presence of root contact in 59 patients (mean age: 21.2, age range: 13-54 years) who had miniscrew implants. Next, 38 miniscrews (1.6 mm in diameter and 8.0 mm in length) with root contact were investigated in the patients. Root contact length was measured using CBCT taken after miniscrew implant placement. RESULTS: The results showed that root contact was the most critical risk factor for failure of miniscrew implants in patients. Logistic regression analysis showed that the risk approximately doubled as the root contact length increased by 1.0 mm. CONCLUSION: Avoiding proximity to the root during miniscrew implant placement is essential to prevent implant failure.
23
Fulya Özdemir, Murat Tozlu, Derya Germeç-Çakan · American Journal of Orthodontics and Dentofacial Orthopedics · 2013 · 159 citations
Joorok Park, Heon Jae Cho · American Journal of Orthodontics and Dentofacial Orthopedics · 2009 · 153 citations