Technology in Cancer Research & Treatment · 2022 · 21 citations · 14 references
MedicineOncologyRadiation OncologyClinical Radiation OncologyRadiobiologyTreatment PlanningExternal Beam RadiotherapyImage-guided RadiotherapyRadiation TherapyRadiation Treatment PlanningTreatment Planning Dose CalculationTreatment Planning PhysicsBiologically Guided PlanningRadiotherapy Dose PlanningTreatment Plan EvaluationTreatment Plan OptimizationTreatment Planning Optimization
Purpose: The first clinical biology-guided radiation therapy (BgRT) system—RefleXion TM X1—was installed and commissioned for clinical use at our institution. This study aimed at evaluating the treatment plan quality and delivery efficiency for IMRT/SBRT cases without PET guidance. Methods: A total of 42 patient plans across 6 cancer sites (conventionally fractionated lung, head, and neck, anus, prostate, brain, and lung SBRT) planned with the Eclipse TM treatment planning system (TPS) and treated with either a TrueBeam ® or Trilogy ® were selected for this retrospective study. For each Eclipse VMAT plan, 2 corresponding plans were generated on the X1 TPS with 10 mm jaws (X1-10mm) and 20 mm jaws (X1-20mm) using our institutional planning constraints. All clinically relevant metrics in this study, including PTV D95%, PTV D2%, Conformity Index (CI), R50, organs-at-risk (OAR) constraints, and beam-on time were analyzed and compared between 126 VMAT and RefleXion plans using paired t-tests. Results: All but 3 planning metrics were either equivalent or superior for the X1-10mm plans as compared to the Eclipse VMAT plans across all planning sites investigated. The Eclipse VMAT and X1-10mm plans generally achieved superior plan quality and sharper dose fall-off superior/inferior to targets as compared to the X1-20mm plans, however, the X1-20mm plans were still considered acceptable for treatment. On average, the required beam-on time increased by a factor of 1.6 across all sites for X1-10mm compared to X1-20mm plans. Conclusions: Clinically acceptable IMRT/SBRT treatment plans were generated with the X1 TPS for both the 10 mm and 20 mm jaw settings.
14
Stereotactic body radiation therapy: The report of AAPM Task Group 101
Stanley Benedict, Kamil M. Yenice, D Followill et al. · Medical Physics · 2010 · 2.2K citations · Full text
Assessment of <sup>68</sup>Ga-PSMA-11 PET Accuracy in Localizing Recurrent Prostate Cancer
Wolfgang P. Fendler, Jérémie Calais, Matthias Eiber et al. · JAMA Oncology · 2019 · 787 citations · Full text
Medicine, Nuclear Medicine, Oncology +17