Silent myelin-weighted magnetic resonance imaging

Tobias Wood, Nikou L. Damestani, Andrew J. Lawrence, Emil Ljungberg, Gareth J. Barker, Ana Beatriz Solana, Florian Wiesinger, Steven Williams

Wellcome Open Research · 2020 · 12 citations · 45 references

DOIFull text

Open access

Concepts

Abstract

<ns4:p> <ns4:bold>Background:</ns4:bold> Inhomogeneous Magnetization Transfer (ihMT) is an emerging, uniquely myelin-specific magnetic resonance imaging (MRI) contrast. Current ihMT acquisitions utilise fast Gradient Echo sequences which are among the most acoustically noisy MRI sequences, reducing patient comfort during acquisition. We sought to address this by modifying a near silent MRI sequence to include ihMT contrast. </ns4:p> <ns4:p> <ns4:bold>Methods: </ns4:bold> A Magnetization Transfer preparation module was incorporated into a radial Zero Echo-Time sequence. Repeatability of the ihMT ratio and inverse ihMT ratio were assessed in a cohort of healthy subjects. We also investigated how head orientation affects ihMT across subjects, as a previous study in a single subject suggests this as a potential confound. </ns4:p> <ns4:p> <ns4:bold>Results:</ns4:bold> We demonstrated that ihMT ratios comparable to existing, acoustically loud, implementations could be obtained with the silent sequence. We observed that the ihMT ratio varied with the orientation of the head. </ns4:p> <ns4:p> <ns4:bold>Conclusions: </ns4:bold> Silent ihMT imaging is a comparable alternative to conventional, noisy, alternatives. For all future ihMT studies we recommend careful attention should be paid to subject positioning within the scanner. </ns4:p>

References

45