Design, Automation & Test in Europe Conference & Exhibition (DATE), 2013 · 2013 · 13 citations · 0 references
EngineeringBiomedical EngineeringElectrical EngineeringAnalog ElectronicsBioinstrumentationNeural EngineeringBiomedical InstrumentationImplantable Biomedical DevicesImplantable MicrosystemsImplantable Neural DevicesNeural Signal AcquisitionPhysiological Signal AcquisitionElectrocorticographyElectrophysiological Signal AcquisitionGold ElectrodesInductive CouplingNeural RecordingNeural Recording SystemsNeural Signal RecordingPower Consumption
This paper presents a fully implantable neural recording system for the simultaneous recording of 128 channels. The electrocorticography (ECoG) signals are sensed with 128 gold electrodes embedded in a 10 µm thick polyimide foil. The signals are picked up by eight amplifier array ICs and digitized with a resolution of 16 bit at 10 kHz. The digitized measurement data is processed in a reconfigurable digital ASIC, which is fabricated in a 0.35 µm CMOS technology and occupies an area of 2.8×2.8mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> . After data reduction, the measurement data is fed into a transceiver IC, which transmits the data with up to 495 kbit/s to a base station, using an RF loop antenna on a flexible PCB. The power consumption of 84mW is delivered via inductive coupling from the base station.