Journal of Applied Clinical Medical Physics · 2003 · 18 citations · 0 references
MedicineRadiation OncologyRadiologyRadiation DosimetryRadiation TransportTreatment PlanningDose CalculationMonte Carlo DosimetryPhantom DosimetryRadiation Treatment PlanningTreatment Planning Dose CalculationTreatment Planning PhysicsComputerized Treatment PlanningDose Distribution CalculationsElectron Density DistributionsMonte Carlo Dose ModelingMonte Carlo Dose SimulationRadiotherapy Dose PlanningTreatment Planning DosimetryTreatment Planning System Dose Verification
The dose calculation accuracy of a commercial pencil beam IMRT planning system is evaluated by comparison with Monte Carlo calculations and measurements in an anthropomorphic phantom. The target volume is in the right lung and mediastinum and thus significant tissue inhomogeneities are present. The Monte Carlo code is an adaptation of the msnp code and the measurements were made with TLD and film. Both the Monte Carlo code and the measurements show very good agreement with the treatment planning system except in regions where the dose is high and the electron density is low. In these regions the commercial system shows doses up to 10% higher than Monte Carlo and film. The average calculated dose for the CTV is 5% higher with the commercial system as compared to Monte Carlo. PACS number(s): 87.53.-j, 87.66.-a