Applied Optics · 1978 · 154 citations · 2 references
MedicineOptometryVision ScienceClinical OptometryEye MovementsOpticsEye Movement AnalysisOcular BiometryEye Movement Data QualityEye Movement MonitoringEye Tracking SensorsGaze TrackingInfrared OculographyOcular Biometrics
A combined optometer–eyetracker instrument measures dynamic refractive power and 3‑D gaze direction of the same eye in real time using near‑IR light, with a −4 to +12 D range and 0.1 D precision, and can be set side‑by‑side for binocular tracking without attaching anything to the subject. The device achieves a >20° field of view, measuring horizontal and vertical gaze with ~1′ noise and repeatability, and produces 3‑D monocular and binocular eye‑movement recordings.
A combined optometer and eyetracking instrument has been developed to measure both the dynamic refractive power and the direction of gaze of the same eye. In effect, this instrument measures, as a function of time, the point in 3-D space on which the eye is fixated. Nothing is attached to the subject (patient), who is easily aligned in the device. The measuring wavelength is in the near ir and is invisible. The usable field of the instrument is greater than 20°; the horizontal and vertical directions of gaze are measured with a noise level and repeatability of about 1 min of arc. The range of the optometer is approximately −4 to +12 diopters; refractive power is measured to about 0.1 diopter. Two instruments may be aligned side by side for tracking both eyes simultaneously. Three-dimensional monocular and binocular eye movement records are shown.
2
Servo-Controlled Infrared Optometer*
Tom N. Cornsweet, Hewitt D. Crane · Journal of the Optical Society of America · 1970 · 115 citations · Full text