Plastic Surgery · 2014 · 11 citations · 3 references
MedicinePlastic SurgerySurgeryBurn SurgeryUrogenital SurgeryUrologic SurgeryBurn CareBurn ManagementBurn ComplicationsElectrocautery TreatmentOperative DebridementPostoperative DebridementRemovalSplit-thickness Skin Graft ReconstructionSplit-thickness Skin Grafts
Ignition of chlorhexidine by an electrocautery unit is rare but can have devastating consequences for the patient and the surgeon. A case involving a 77-year-old man who underwent removal of an indwelling artificial urethral sphincter is presented. The chlorhexidine was ignited when the urologist activated the electrocautery unit, causing third-degree burns to the patient. A plastic surgeon treated the burns with surgical debridement and split-thickness skin grafting. A systematic review of the literature was performed with best practice recommendations. To the authors' knowledge, the present case is the ninth such case reported.
3
Inflammable antiseptics and theatre fires
Charlton Briscoe, Dennis W. Hill, J. P. Payne · British journal of surgery · 1976 · 42 citations
Burn Center Management of Operating Room Fire Injuries
Linwood R. Haith, Wil Santavasi, Tyler K. Shapiro et al. · Journal of Burn Care & Research · 2012 · 25 citations
Medicine, Surgery, Burn Surgery +8