Continuous flow method for the determination of phosphorus using the fluorescence quenching of Rhodamine 6 G with molybdophosphate

Shoji Motomizu, Hiroshi Mikasa, Mitsukom OSHIMA, Kyoji Tôei

BUNSEKI KAGAKU · 1984 · 15 citations · 0 references

DOIFull text

Open access

Concepts

Abstract

Continuous flow method was examined for the determination of phosphorus. The molybdophosphate formed between orthophosphate and molybdate in hydrochloric acid medium diminished the fluorescence of Rhodamine 6 G. Carrier solution (distilled water) and reagent solution were propelled by double plunger pump P1 and P2 (flow rate : 0.98 ml/min), and sample solution(160μl) was injected into the carrier stream. The two streams were mixed in 20 cm long Teflon tubing (1 mm i.d.), and the mixture was flowed through a flow cell(18μl), at which the fluoresence of Rhodamine 6 G was detected(λex =350 nm, λem=580 nm). The reagent solution consists of 0.035 M molybdenum and 1 × 10-5 M Rhodamine 6 G in 0.8 M hydrochloric acid. Co-existing ions generally existing in river and sea waters did not interfere the determination of phosphorus. The calibration curve was linear from 0 to 45 ppb of phosphorus. The method was applied to sea water. The sampling rate was 20 samples per hour.