Prospective acquisition correction for head motion with image‐based tracking for real‐time fMRI

Stefan Thesen, O. Heid, Edgar A. Mueller, Lothar R. Schad

Magnetic Resonance in Medicine · 2000 · 24 citations · 10 references

Concepts

Abstract

In functional magnetic resonance imaging (fMRI) head motion can corrupt the signal changes induced by brain activation.This paper describes a novel technique called Prospective Acquisition Cor-rEction (PACE) for reducing motion-induced effects on magnetization history.Full three-dimensional rigid body estimation of head movement is obtained by image-based motion detection to a high level of accuracy.Adjustment of slice position and orientation, as well as regridding of residual volume to volume motion, is performed in real-time during data acquisition.Phantom experiments demonstrate a high level of consistency (translation < 40m; rotation < 0.05°) for detected motion parameters.In vivo experiments were carried out and they showed a significant decrease of variance between successively acquired datasets compared to retrospective correction algorithms.

References

10