Catheter Ablation Without Fulguration: Design and Performance of a New System

Arif Ahsan, David Cunningham, Edward Rowland, A F Rickards

Pacing and Clinical Electrophysiology · 1989 · 21 citations · 4 references

Concepts

Abstract

A new system is described for catheter ablation which comprises a short time capacitive power source and a specially designed catheter. The aim of the new system is to deliver high voltages with minimal or no arcing and thus avoid the risks associated with barotrauma. The performance of the new system was compared in a saline tank with that of the conventional system. The new system allowed significant increases in delivered voltage, current and energy without arcing. This new system should allow safer catheter ablation.

References

4