EngineeringNuclear EngineeringPhysicsNuclear PhysicsRadiation DetectionRadiation MeasurementBackground Radiation MeasurementBackground ReductionDark MatterDetector TechnologyGamma-ray InstrumentationGamma-ray SpectroscopyDark Matter Direct DetectionGamma SpectroscopyGamma-ray SpectrometryLaboratory Background MeasurementLow-background MeasurementNuclear RecoilRadon Background MeasurementRadon Measurement
A bstract PandaX-4T is a ton-scale dark matter direct detection experiment using a dual-phase TPC technique at the China Jinping Underground Laboratory. Various ultra-low background technologies have been developed and applied to material screening for PandaX-4T, including HPGe gamma spectroscopy, ICP-MS, NAA, radon emanation measurement system, krypton assay station, and alpha detection system. Low background materials were selected to assemble the detector. Surface treatment procedures were investigated to further suppress radioactive background. Combining measured results and Monte Carlo simulation, the total material background rates of PandaX-4T in the energy region of 1–25 keV ee are estimated to be (9.9 ± 1.9) × 10 − 3 mDRU for electron recoil and (2.8 ± 0.6) × 10 − 4 mDRU for nuclear recoil. In addition, nat Kr in the detector is estimated to be < 8 ppt.
17
Particle dark matter: evidence, candidates and constraints
Gianfranco Bertone, Dan Hooper, Joseph Silk · Physics Reports · 2004 · 4.9K citations · Full text
Current status of direct dark matter detection experiments
Jianglai Liu, Xun Chen, Xiangdong Ji · Nature Physics · 2017 · 255 citations · Full text
Material radioassay and selection for the XENON1T dark matter experiment
E. Aprile, J. Aalbers, F. Agostini et al. · The European Physical Journal C · 2017 · 65 citations · Full text