Publication | Open Access
Scanning Distance Influence on the Intraoral Scanning Accuracy—An In Vitro Study
46
Citations
11
References
2022
Year
Intraoral ScannerEngineeringMicroscopyMeasurementBiometricsAccuracy And PrecisionOral MedicineDigital ImpressionBiomedical EngineeringClinical DentistryDental RadiologyDistance InfluenceDigital ImpressionsIntraoral Scanning Accuracy—anRadiologyHealth SciencesOphthalmologyMedical ImagingDigital ImagingRadiographic ImagingDental ImagingIntraoral ScannersBiomedical ImagingVitro StudyDental HygieneOrthodontic Research
Intraoral scanners (IOS) have reached a point where their impact in the dental office cannot be denied. The distance between the tip of the IOS and the preparation may have implications on the accuracy of the digital model. The objective of this study was to evaluate the differences in accuracy between digital impressions in the scenario of different scanning distances. Twenty consecutive scans were performed at five predetermined distances: 5 mm, 10 mm, 15 mm, 20 mm and 23 mm by a single operator. The scanning distance of 10 mm displayed the best accuracy with an overall trueness value of 23.05 μm and precision value of 4.2 μm. The drawn conclusion was that increased scanning distances can decrease the accuracy of a digital impression.