Simultaneous determination of vitamin D metabolites 25(OH)D3 and 1α,25(OH)2D3 in human plasma using liquid chromatography tandem mass spectrometry
17
Citations
24
References
2022
Year
The method was acceptably linear over the concentration ranges of 20-1200 pg/mL for 1α,25(OH)<sub>2</sub>D<sub>3</sub> and 1-60 ng/mL for 25(OH)D<sub>3</sub>, respectively, with correlation coefficients of r<sup>2</sup> > 0.993. Both the inter-assay and intra-assay precision was < 15%, and the analytical recoveries were within 100% ± 10%, with no significant matrix effect or carryover. Thereby, we, provide a facile method for the simultaneous detection of vitamin D metabolites in plasma.