Medical Imaging 2022: Physics of Medical Imaging · 2022 · 10 citations · 8 references
EngineeringElectrical EngineeringRadiologyPhotonicsRadiation DetectionRadiation MeasurementDetector TechnologyRadiation DetectorsScintillation DetectorsSemiconductor DetectorsX-ray DetectionX-ray DetectorsDetectorsDirect Conversion DetectorsPhotomultiplier DetectorsPhoton CountingPhoton-counting DetectorsPixel DetectorsSilicon DetectorsSilicon Photomultipliers
We investigate X-ray photon-counting scintillation detectors with silicon photomultiplier (SiPM) readout. These circumvent some drawbacks of direct-conversion detectors. We measured observed count rate (OCR) versus X-ray tube current for single-pixel detectors consisting of LYSO:Ce and YAP:Ce scintillators coupled to ultrafast SiPMs. For a 30 keV threshold, the maximum OCRs equal 4.5 Mcps/pixel (LYSO:Ce) and 5.5 Mcps/pixel (YAP:Ce) for paralyzable-like counting and 10 Mcps/pixel (LYSO:Ce) and 12.5 Mcps/pixel (YAP:Ce) for nonparalyzable-like counting. We estimate that the twice as fast LaBr3:Ce scintillator yields OCRs approaching those of CdTe/CZT-based photon-counting CT detectors. We also show energy response data and discuss dose-efficient pixel miniaturization.
8
S. Kappler, André Henning, Björn Kreisler et al. · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2014 · 77 citations