End to end comparison of surface‐guided imaging versus stereoscopic X‐rays for the SRS treatment of multiple metastases with a single isocenter using 3D anthropomorphic gel phantoms
30
Citations
45
References
2022
Year
Dose metrics for a setup with surface imaging was found to be consistent with setup using x-ray imaging, demonstrating high accuracy and reproducibility for treatment delivery. Results indicate the feasibility of using surface imaging for position verification at noncoplanar couch angles for single-isocenter, multiple-target SRS using end-to-end quality assurance (QA) testing with 3D polymer gel dosimetry.