MedicineOncologyRadiation OncologyRadiation DosimetryRadiation MeasurementRadiation PhysicsElectron DosimetryPhoton DosimetryRadiation DetectorsRadiation Dose MeasurementRadiation TherapyX-ray DetectorsClinical Photon Beam DosimetryDetectorsExternal Beam Radiotherapy DosimetryIon Recombination CorrectionsSolid State DosimetrySolid-state DetectorsStereotactic Radiosurgery DosimetryX-ray Dosimetry
INTRODUCTION: Kilovoltage (kV) X-ray beams are an essential modality in radiotherapy. Solid state detectors are widely available in radiotherapy departments, but their use for kV dosimetry has been limited to date. This study aimed to evaluate the dosimetric performance of a range of solid state detectors for kV dosimetry. METHOD: . The following Physikalisch-Technische Werkstätten (PTW) dosimeters were used for measurements: Advanced Markus, PinPoint 3D and Semiflex ionization chambers; photon, electron, and stereotactic radiosurgery (SRS) diodes plus the microDiamond detector. All PDDs were normalized at 5 mm depth, and ROFs were measured at 3 mm depth to avoid collisions with the end of the applicators. ROFs measured using chambers were corrected for polarity and ion-recombination effects. RESULTS AND DISCUSSION: PDD measurements for 60, 100, and 150 kVp beams exhibited good agreement between all diodes and the ionization chambers over the entire range of depths except in the first few millimeters near the surface. However, for the 300 kVp, all diode detectors exhibited an overresponding behavior compared to reference depth dose data measured with the Advanced Markus chamber. ROFs with the diodes were higher than the Advanced Markus chamber at low energy, and the magnitude of these differences is inversely proportional to the field sizes. The PTW P diode showed the highest variation of up to 15% in the output factor compared to the Advanced Markus chamber. CONCLUSION: This study evaluated the dosimetric performance of a range of solid state detectors in kV relative dosimetry. This study showed that diode detectors are a suitable replacement for ionization chambers for the PDD measurement of low energy kV beams (60-150 kVp) except for the PDD of 60 kVp with the smaller field sizes. However, an overresponding behavior of diode detectors at 300 kVp beams shows that diode detectors are not suitable for the PDD measurement of high energy kV beams. Generally, all solid state detectors overresponded to ROF measurements, indicating that it is not suitable for ROF measurements. In general, both shielded and unshielded diodes produced a similar dosimetric response, which demonstrates that the energy dependence of solid state detectors should be considered before they are used for any kV relative dosimetric measurements.
21
AAPM protocol for 40–300 kV x‐ray beam dosimetry in radiotherapy and radiobiology
Chao Ma, C Coffey, L DeWerd et al. · Medical Physics · 2001 · 819 citations · Full text
Radiotherapy for epithelial skin cancer
Jay Locke, Shervin Karimpour, Gwen Young et al. · International Journal of Radiation Oncology*Biology*Physics · 2001 · 272 citations · Full text
Prepared by a Working Party of the, R J Aukett, R. M. Harrison et al. · Physics in Medicine and Biology · 1996 · 258 citations
Radiation Oncology Physics: A Handbook for Teachers and Students
K. Kainz · Medical Physics · 2006 · 216 citations