Real-life performance of a COVID-19 rapid antigen detection test targeting the SARS-CoV-2 nucleoprotein for diagnosis of COVID-19 due to the Omicron variant

Paula de Michelena, Ignacio Torres, Ángela Ramos-García, Victoria Gosalbes, Nidia Ruiz, Ana Sanmartín, Pilar Botija, Sandrine Poujois, Dixie Huntley, Eliseo Albert,

medRxiv · 2022 · 14 citations · 6 references

DOIFull text

Open access

Concepts

Abstract

ABSTRACT Objectives It has been suggested that rapid antigen detection assays (RADT) may perform suboptimally in terms of sensitivity for the diagnosis of SARS-CoV-2 Omicron variant infection. To address this issue, we conducted a prospective study in primary health centers to evaluate the clinical performance of the Panbio™ COVID-19 Ag Rapid Test Device in nasopharyngeal specimens (NP) carried out at the point of care. Methods We recruited 244 patients (median age, 40 years; range 2–96; 141 female) with clinical suspicion of COVID-19 (232 adults and 12 children). 228/244 patients had been fully vaccinated (two doses) with licensed COVID-19 vaccines prior to recruitment. Most patients (222/244) were SARS-CoV-2 naïve prior to enrollment. Patients were tested by RT-PCR and RADT within 5 days since symptoms onset. Results 126 patients (51.6%) tested positive by both RT-PCR and RADT, 90 patients (36.8%) returned negative results by both assays and 28 patients (11.4%) yielded discordant results (RT-PCR+/RADT-). No patients tested RT-PCR-/RADT+. Overall specificity and sensitivity of RADT was 100% (95% CI, 95.9–100%) and 81.8% (95% CI, 75–87.1%) respectively. The sensitivity of the assay increased from 79.6% (95% CI, 66.4–88.5) when considering specimens collected at days 0–1 after symptoms onset, to 86.4% (95% CI, 66.7–95.3) when grouping the specimens obtained on days 4–5. Conclusion The Panbio™ COVID-19 Ag Rapid Test Device perform well (≥80% sensitivity) as a point-of-care test for early diagnosis of COVID-19 due to the Omicron variant in primary healthcare centers.

References

6