JASA Express Letters · 2022 · 13 citations · 17 references
Veterinary MedicineBiomedical EngineeringRadiologyDiagnostic ImagingUltrasonographyUltrasoundDiagnostic UltrasonographyMedical ImagingUltrasound ImagingCompoundingLow-contrast DetectabilitySignal-to-noise Ratio ImprovementSpeckle Noise ReductionSpeckle ReductionSynthetic Aperture ImagingUltrasound Lesion Assessment
Speckle variance in ultrasound images limits the detection of low-contrast targets. In conventional compounding, multiple correlated sub-images are generated and then averaged to reduce the speckles at the cost of resolution loss. In this paper, a decorrelation procedure was applied to the correlated sub-images to further reduce speckle variance. Lesion signal-to-noise-ratio (lSNR), which combines the effect of speckle reduction and resolution loss, was used as an indicator of the detectability of lesions. The lSNR of the hyperechoic lesion in the simulated and experimental images using decorrelated compounding was increased by 122% and 89%, respectively, compared to the delay-and-sum method.
17