Commissioning and routine quality assurance of the Radixact Synchrony system
13
Citations
22
References
2021
Year
These results highlight the large improvements in the dose distribution when motion is accounted for with the Synchrony system as measured with a range of phantoms and motion platforms that the majority of users will have available. The testing methods and QA procedures described provide guidance for new users of the Radixact Synchrony system as they implement their own QA programs for this system, until such time as an AAPM task group report is made available. QA procedures including Kilovolts (kV) imaging quality metrics and imaging dose parameters, dose deposition accuracy, target detection coincidence, and target position detection accuracy are described.