An automated chemiluminescent immunoassay (CLIA) detects SARS-CoV-2 neutralizing antibody levels in COVID-19 patients and vaccinees
32
Citations
15
References
2021
Year
With the RBD-based CLIA protocol, CLIA can be used to replace MNT to test SARS-CoV-2 NAb. Vaccine effectiveness, protectiveness and durability can be evaluated effectively by mathematical models. It is RESULTS: Analysing the relationship between NAb titers and concentrations, R<sup>2</sup> for the decision-making tree was 0.870 and that of progressive linear fitting was 0.821. The receiver operating characteristic curve indicated specificity of 78.1%, sensitivity of 87.4%, cut-off value of 6.43 AU/mL and borderline range of 5.79-7.07 AU/mL for CLIA. Three-quarters (75.4%) of vaccinees were found to be NAb positive, and 5.35% vaccinees had NAb protective capability. The half-life of NAb in vaccinees was 10-11 weeks.for vaccinees to take a NAb assay periodically.