Simple 13C-urea breath test with infra-red spectrophotometer.

16

Citations

0

References

1996

Year

Abstract

When mass spectrophotometric analysis is used for the 13C-urea breath test to assess H. pylori infection, it is costly, complicated, and time-consuming. To overcome these disadvantages, we utilized an infra-red spectrophotometer as a substitute for the mass spectrophotometer. A total of 153 patients (181 tests) analyzed with peptic ulcers or non-ulcer dyspepsia were investigated. Breath samples were collected 15 min after ingestion of 13C-urea (100 mg in 30 ml water). An infra-red spectrophotometer was used to determine the concentration of 13CO2 in the expirate. The 13CO2/12CO2 ratio was also measured by mass spectrophotometry to compare results with those of infra-red spectrophotometric analysis. Direct detection of H. pylori was qualified in biopsy specimens. Of the 181 biopsies, 138 were positive for H. pylori infection and 43 were negative. With the urea breath test, the mean value in the positive group was significantly higher than that in the negative group (0.062 +/- 0.044 vs 0.011 +/- 0.014, respectively). The cut-off level, 0.01, was determined as delta 13C atom %. The sensitivity of infra-red spectrophotometry was 97.8% (135/138) and specificity was 74.4% (32/43). There was an extremely high coefficient of correlation (r = 0.996) between mass and infra-red photometric analysis. Infra-red spectrometry appears to have great potential not only for diagnosing H. pylori infection but also for assessing treatment results. Its advantages include technical simplicity, cost-effectiveness, and high accuracy.