Accuracy of noninvasive oxyhemoglobin saturation, end-tidal carbon dioxide concentration, and blood pressure monitoring during experimentally induced hypoxemia, hypotension, or hypertension in anesthetized dogs
66
Citations
11
References
1998
Year
The pulse oximeters tested provided an accurate estimation of SaO2 at SpO2 > 70%. A PEtCO2 value > 55 mm of Hg may represent hypercapnia that is more profound than indicated. Systolic BP determinations were most accurate during hypotensive states and least accurate during hypertension. Diastolic BP measurements were generally more accurate during hypertension than normotension. Accuracy is not appreciably affected by hypotension resulting from vasodilation or blood loss. The tendency to underestimate systemic arterial BP should not interfere with trend detection during unstable clinical conditions.