PubMed · 1985 · 10 citations · 0 references
MedicineDermatologyPhysiologyClinical OptometryColor VisionPhotodermatologyHemodynamicsOxygen TherapyArterial Oxygen SaturationAssessment Of Oxygen SaturationOxygen SaturationOxygen Saturation MeasurementOxygen Saturation MonitoringPulse OximetersSkin Pigmentation DisordersSkin Pigmentation Effects
Noninvasive oximetry has been shown to be a reliable and convenient way of estimating arterial oxygen saturation. Studies have shown that skin pigmentation did not affect the accuracy of oximetry. However, these studies were performed on healthy subjects and the number studied was small. Our study on 31 pigmented patients in the clinical situation showed that the Biox III oximeter gave readings that closely approximated oxygen saturation measurements from arterial blood (y = 4.75 + 0.93x, r = 0.91, p less than 0.001), confirming the usefulness of noninvasive oximetry even in pigmented patients.