Anaesthesia and Intensive Care · 2011 · 59 citations · 9 references
Health ProfessionsRespiratory TherapyPulmonologyResuscitationVentilator ManagementAirway SupportBreathing SupportCardiopulmonary ResuscitationCardiopulmonary Resuscitation For DrowningMechanical VentilationRespiratory ComplicationsCompressionCritical Care After DrowningSupraglottic Airway Effects
We report the failure of an i-gel and an Ambu AuraOnce supraglottic airway to ventilate a drowning victim. Failure was attributed to changes in lung physiology following submersion and inhalation of water that may have required ventilation pressures up to 40 cmH2O to treat the victim's hypoxaemia. The ease of use and rapid insertion of supraglottic airways without interrupting cardiac compression has prompted recommendations for their use during resuscitation. The relatively low leak pressures attainable from many supraglottic airways, however may cause inadequate lung ventilation and entrainment of air into the stomach when these devices are used in drowning victims.
9
Jasmeet Soar, Gavin D. Perkins, Gamal Abbas et al. · Resuscitation · 2010 · 603 citations
THE PRESSURE-VOLUME DIAGRAM OF THE THORAX AND LUNG
Hermann Rahn, Arthur B. Otis, Leigh E. Chadwick et al. · American Journal of Physiology-Legacy Content · 1946 · 530 citations
Medicine, Anatomy, Physiology +12