EngineeringBioengineeringBiomedical EngineeringAnalytical ChemistryBiosensorsSensorsBlood Glucose MonitoringContinuous Glucose MonitoringGlycemic MonitoringNanomaterial-based BiosensorsOptical BiosensingOptical BiosensorsCadmium TellurideFluorescence SensingGlucose BiosensorsGlucose ConcentrationGlucose MonitoringGlucose OxidaseGlucose SensingGlucose Sensors
), which quenched the red fluorescence of CdTe QDs. The fluorescence images of the sensor were obtained using a CCD camera, and the linear relationship between the glucose concentration and the gray value of the fluorescence image was established. The analyzer shows good sensitivity (LOD, 0.12 mM) with a wide linear range of 0.12-26 mM. Based on the linear relation, the software of the analyzer was written in the C++ language, which can automatically give the gray value of the image and the corresponding glucose concentration. The UG analyzer was used for the detection of a large number clinical samples and compared with a variety of UG test papers, which all showed good detection performance. The novel analyzer we proposed has an important significance in the screening of diabetes and the self-management of diabetic patients.
34