NeuroImage · 2021 · 90 citations · 38 references
EngineeringBiomedical EngineeringNeuroscienceBiomedical ImagingDiagnostic ImagingMetabolismBrain Energy MetabolismBrain ImagingMagnetic Resonance ImagingMagnetic Resonance SpectroscopyMedical ImagingMedical Imaging InstrumentationChemical Shift ImagingCoil DesignGlucose Metabolism ImagingMagnetic Field StrengthMetabolic Brain ImagingMetabolic ImagingMetabolite ImagingRegional Brain Metabolism
To present first highly spatially resolved deuterium metabolic imaging (DMI) measurements of the human brain acquired with a dedicated coil design and a fast chemical shift imaging (CSI) sequence at an ultrahigh field strength of B0 = 9.4 T. 2H metabolic measurements with a temporal resolution of 10 min enabled the investigation of the glucose metabolism in healthy human subjects. The study was performed with a double-tuned coil with 10 TxRx channels for 1H and 8TxRx/2Rx channels for 2H and an Ernst angle 3D CSI sequence with a nominal spatial resolution of 2.97 ml and a temporal resolution of 10 min. The metabolism of [6,6′-2H2]-labeled glucose due to the TCA cycle could be made visible in high resolution metabolite images of deuterated water, glucose and Glx over the entire human brain. X-nuclei MRSI as DMI can highly benefit from ultrahigh field strength enabling higher temporal and spatial resolutions.
38
Deuterium metabolic imaging (DMI) for MRI-based 3D mapping of metabolism in vivo
Henk M. De Feyter, Kevin L. Behar, Zachary Corbin et al. · Science Advances · 2018 · 374 citations · Full text