Publication | Open Access
Clinical validation and evaluation of a novel six-lead handheld electrocardiogram recorder compared to the 12-lead electrocardiogram in unselected cardiology patients (EVALECG Cardio)
35
Citations
10
References
2021
Year
Several parameters recorded by the Kardia 6L (QT interval in all six leads, rhythm detection, PR interval, QRS duration, and cardiac axis) perform closely to the gold standard 12-lead ECG. However, that consistency weakens for left ventricular hypertrophy, QRS amplitudes (Lead I and AVL), and ischaemic changes.