Diagnosis of von Willebrand disease: An assessment of the quality of testing in North American laboratories
17
Citations
8
References
2021
Year
Imprecision was lowest for VWF antigen and highest for collagen binding activity (CBA) with median coefficient of variation (CV) of 12% (interquartile range (IQR) 7%) and 23% (IQR 21%) respectively. Within the VWF PBA assays, the gain-of-function mutant GP1b binding (VWF: GP1bM) methods had the least imprecision (CV 9%, IQR 10%). All assays, including the various PBA methods had excellent consensus. The majority of laboratories agreed that normal (median consensus-82%, IQR 16%) and T1 VWD (median consensus-100%, IQR 9%) samples had normal multimer distribution. Consensus among laboratories for final interpretations was excellent for normal samples (median 81%, IQR 8%), good for T1 VWD samples (median 59%, IQR 9%), and fair for T2 VWD samples (median 44%, IQR 21%). Consensus on final interpretation decreased as sample complexity increased.