Publication | Open Access
Multicenter Evaluation Study Comparing a New Factory-Calibrated Real-Time Continuous Glucose Monitoring System to Existing Flash Glucose Monitoring System
25
Citations
8
References
2021
Year
Real-time MonitoringDiabetes ManagementMeasurementCalibrationInsulin ManagementDiabetesLaboratory MedicineEducationBlood Glucose MonitoringMulticenter Evaluation StudyAidex Cgm SystemInstrumentationInsulin DeliveryMedicineArtificial PancreasReal-time Glucose MonitoringAbbott Libre
This study reports a clinical evaluation of AiDEX CGM system featuring a 14-day sensor, real-time glucose monitoring and factory-calibration. A multicenter, prospective, masked clinical study was conducted at with a total of 120 participants. Each participant wore 4 studied sensors and had one in-clinic visit for venous blood reference tests. 40 out of the 120 participants wore additional Abbott Libre sensors and performed at least 7 capillary BG tests daily for additional reference and comparison. Continuous glucose error grid analysis (CG-EGA) showed that AiDEX and Abbott Libre had good agreement with venous blood glucose, with 98.69% and 98.96% accurate readings, respectively. Overall MARD of AiDEX CGM systems was 9.08% when compared to venous blood reference and 10.1% when compared to finger capillary BG reference.